Model Number |
Miracle NC8107 |
A. Machine
- Miracle8107 MACHINE
- RENISHAW UCC Lite Control System
- Calibrate Ball with Support Base
- User and System Manuals
B. Computer System
- Minimum configuration: Pentium Ⅳ3.0 GHz/512MB RAM/80GB Hard Disk/32MB Graphic Card /17″Color LCD Monitor/ Window’s OS
- HP Color DeskJet printer
C. Software System
-Amereican Rantional DMIS software(for Windows XP)
D. RENISHAW Probe System 1Set
- MH20i Renishaw Motorized Probe Head
- Renishaw Probe Styli Kit
E. Installation at Customer Site
System installation includes start-up, adjust and metrology test. The metrology test will be performed according the (GB/T16857.2-1997)standard regulation and it equal to ISO 10360-2.
F. Application Training
Training on the CMM operation & application 5 working days at The User’s site.
G. Warranty
12 Months guarantee free of charge after Final Acceptance.
Miracle 8107 Performance Specifications
Measuring Strokes : 800(mm) x1000(mm) x 700(mm)
Overall Dimension : 1700(mm) x 2150(mm) x3000 (mm)
Max part weight : 1000 kg
Resolution : 0.2 (mm)
Accuracy according GB/T16857.2-1997 equal ISO 10 360:
(Volumetric length measuring error) E =2.9 +4L/1000 mm
(Volumetric Probing Error) R = 3.0 mm
Environment Specifications
Room required specifications
- Average room temperature: 18 - 22 °C
- Maximum thermal gradients:1 °C/hour,2 °C/day,1 °C/m,
- Air supply system
- (Min. air pressure) : 0.7 MPa
- (Min. air consumption) : 0.1 m3/min
Electrical requirements
- (Voltage) : 220 Vac ± 10%
- (Frequency) : 50 Hz
- (Current) : 10 A
Allowed Maximum Floor Vibration
- (Frequency Range): 0 ~ 19 Hz 19~75 Hz > 75 Hz
- (Maximum Amplitude): 0.0011 mm 0.0011~0.002 mm 0.002 mm
|